This service note describes how to reset the “catalyst replacement warning” LED on NGC and NGT boards.

The catalytic converter (catcon) has a limited lifespan and therefore needs replacement after 3 years of operation (please see the operation manual). As a reminder to the operator, the generator will display a warning message “Catalyst life alarm”, when the life of the catcon is exceeded. The red external “check system” LED begins to flash. From this time on, the operator has 60 days to replace the catcon. When this time expires, the control board disables the system.